Full Disclosure License Management Setup - GE MP100 Series Service Manual

Clinical information center
Table of Contents

Advertisement

Full disclosure license management setup

Overview
Set the full disclosure defaults
2026419-033E
5.
After making your selections, choose one of the following:
Click OK to apply your changes and close the CIC Setup window.
Click Apply to apply your changes without closing the CIC Setup window.
6.
Verify the new settings prior to monitoring patients.
Every CIC Pro center with v4.0.x or later software has the ability to store 16 beds of
full disclosure data for one hour without a license. A CIC Pro center running v5.0.x or
later software with the full disclosure option has the ability to store up to 16 beds for
72 hours with the appropriate licenses and license settings.
When operating two CIC Pro centers within the same care unit, with one having the
full disclosure option, and the other without: It is necessary to turn full disclosure
OFF on the CIC Pro center without the full disclosure option, to prevent a conflict
between the two different storage capacities of the CIC Pro centers.
To turn off full disclosure on the CIC Pro center without the full disclosure option,
you must set the full disclosure License type to none.
defaults on page 6-47.
This option sets the full disclosure settings. In user mode, only the full disclosure
Report and Strip settings are configurable.You must be in the Service mode to set the
other full disclosure settings at the CIC Pro center.
Complete the following procedure to configure the Full Disclosure Defaults settings:
1.
From the multi-patient viewer, click Setup CIC.
2.
Log on to Setup CIC with service access on page
3.
From the CIC Setup screen, click Full Disclosure Defaults.
CIC Pro™
Configuration
See Set the full disclosure
4-5.
6-47