Checkout Full Test - GE Carestation 750 Clinical Reference Manual

Anesthesia delivery system
Hide thumbs Also See for Carestation 750:
Table of Contents

Advertisement

Checkout full test

2
1
The full test runs automatically and beeps to indicate when it is finished or if interaction is required.
Perform a Full Test when any component of the system is changed (breathing system, vaporizers, pipeline inlets).
The Full Test does the following tests: Vent and Gas, Vaporizer Leak, and Circuit Leak. When one of the tests is
completed, the next test begins.
1. Select Checkout.
2. From the Checkout menu, select Full Test and follow the instructions.
If a test fails, follow the instructions to perform a retest or accept the results, or select Help to follow the
troubleshooting tips to check the problem.
Note:
In case of a patient emergency, the Full Test may be bypassed by selecting Start Case and then selecting
Bypass. The general message Please Do Checkout is displayed if a Full Test or all of the individual tests are not
completed with passing results within 24 hours. The messages Checkout has not passed and Checkout bypass
will be recorded to the system log and will be displayed in Checkout menu when checkout is bypassed.
21