Visual inspection

Cleaning

2026419-033E
General condition: Carefully inspect the equipment for physical damage to the
case, the display screen, and the controls. Do not use the equipment if damage is
determined. Refer damaged equipment to qualified service personnel.
Connectors: Inspect all external connectors for damaged pins, prongs and
connector housings. Refer damaged equipment to qualified service personnel for
repair. Ensure that all connections are properly seated and secure.
Cable insulation: Inspect all external cables, insulation, strain-reliefs and
connectors for damage, cracks or degradation. Refer damaged equipment to
qualified service personnel for repair.
Display: Inspect the display face for marks, scratches, or other damage. Physical
damage to a touchscreen face may jeopardize its performance. Replace if
necessary.
Fans: Inspect all cooling fans in the CIC Pro center to make sure they are in good
working order and are not clogged with dust. These fans draw outside air into the
unit to cool internal components.
Safety labels and inscriptions: Check that safety labels and inscriptions on the
device are clearly legible.
Mounting hardware: Check for loose or missing screws on the mounting
hardware.
CAUTION
Never immerse the device in any liquid.
CAUTION
Do not pour or spray any liquid directly on the device or permit fluid
to seep into connections or openings.
CAUTION
Never use conductive solutions, solutions that contain chlorides,
wax, or wax compounds to clean the device.
CAUTION
Never autoclave or steam clean the device.
CIC Pro™