Electromagnetic compatibility

Electromagnetic compatibility (EMC)

Guidance and manufacturer's declaration

Electromagnetic emissions

Emissions test
RF Emissions (radiated)
EN 55011
A-2
Changes or modifications to this system not expressly approved by GE could cause
EMC issues with this or other equipment. This system is designed and tested to
comply with applicable regulation regarding EMC and needs to be installed and put
into service according to the EMC information stated in this appendix.
CAUTION
EMC — Magnetic and electrical fields are capable of interfering
with the proper performance of the device. For this reason make sure
that all external devices operated in the vicinity of the monitor
comply with the relevant EMC requirements. X-ray equipment or
MRI devices are a possible source of interference as they may emit
higher levels of electromagnetic radiation.
CAUTION
Use of portable phones or other radio frequency (RF) emitting
equipment near the system may cause unexpected or adverse
operation.
CAUTION
The equipment or system should not be used adjacent to, or stacked
with, other equipment. If adjacent or stacked use is necessary, the
equipment or system should be tested to verify normal operation in
the configuration in which it is being used.
The CIC Pro center is intended for use in the electromagnetic environment specified
below. It is the responsibility of the customer or user to assure that the CIC Pro center
is used in such an environment.
Compliance
Group 1
The device uses RF energy only for its internal function. Therefore, its RF
Class A
emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
CIC Pro™
Electromagnetic environment – guidance
2026419-033E