Emergency; Partial Ups Fluoro (Option); Full Ups Record (Compatability) - GE Innova 2121-IQ Biplane Preinstallation Manual

Cardiovascular imaging system
Table of Contents

Advertisement

GE Healthcare

12 Emergency

12.1 Partial UPS Fluoro (option)

A Fluoro UPS (20 kVA) has been designed for Innova 2121-IQ, 3131-IQ. This partial UPS lets the
customer complete an exam in fluoro mode in case of a power failure. The autonomy provided by
this UPS is 5 minutes of fluoro every 24 hours.

12.2 Full UPS Record (compatability)

An UPS sized for a minimum of 150 kVA is required to supply the system in record mode. Such
an UPS would provide to the customer about 10 minutes of autonomy.
The UPS output shall follow the same output specifications as shown in
transformer specifications in
Contact your local GE Healthcare representative for more information and/or public power
distribution analysis.