5. Hygienic preparation

Valve cage with
expiration cone
Patient valve
Safety valve
(relief valve)
Squeeze
bag
Intake port inner
part with
diaphragm
Intake port cover
with holding ring
COMBIBAG should undergo hygienic purification af-
ter every use.
Upon completion of the purification process, carry out
a functional test.
Individual parts of the COMBIBAG resuscitator are
made of silicon or polysufone.
Use soapy water to
clean
running water. For cleaning purposes, take apart the
bag, separating the pieces into components as de-
scribed.
When assembling or disassembling
Warning:
COMBIBAG, do not use any sharp-edged objects.
The patient valve does not have to be dismantled in
order to be cleaned!
Should you choose to dismantle the valve anyway,
not unscrew the cap of the valve
the pressure setting of the valve.
To
the equipment, use a disinfectant that is suit-
disinfect
able for rubber and polysulfone. Be sure to observe the
instructions for the disinfectant. We recommend
GIGASEPT FF.
Sterilization with steam, hot air or gas should be car-
ried out according to standard practices after careful
cleaning of the individual parts.
The oxygen reservoir
cannot be sterilized
cleaned and disinfected.
COMBIBAG and rinse under
since that could change
. It should be
Hygienic preparation