5.1 Cleaning during use

5.2 Maintenance and service

16
Hygienic preparation
Should the valve become blocked as a result of regur-
gitation, clean it as follows:
Unscrew the valve cage.
Clean the valve element and valve body with the
finger or by gentle tapping.
Rapidly squeeze the bag several times to blow out
any foreign particles.
Screw in the valve cage, check the function and
continue ventilation.
Always check the function of the resuscitator after
Note:
cleaning and re-assembly!
Partially cover the patient connection.
Squeeze the bag and check the valve element for
correct function and leak-tightness (see "4.2 Check-
ing the function" on page 10).
With proper use and maintenance of the COMBIBAG
no additional service is required. Just the intake dia-
phragm should be examined once a year. The dia-
phragm is ready for proper use if no cracking spots are
visible.
Wavy, distorted or sticked items have to be replaced.
A frequent functional check is recommended to be car-
ried out according to Pt. "4.2 Checking the function".