DISPlay Subsystem
ACCumulate
ACCumulate
Command
:DISPlay:ACCumulate {{ON|1}|{OFF|0}}
The ACCumulate command works in conjunction with the commands in the
Acquisition Subsystem. In the Normal mode, the ACCumulate command
turns the infinite persistence on or off.
Example
OUTPUT XXX;":DISPLAY:ACC ON"
Query
:DISPLAY:ACCumulate?
The ACCumulate query reports if accumulate is turned on or off.
Returned Format
[:DISPlay:ACCumulate] {1|0}<NL>
Example
OUTPUT XXX;":DISPLAY:ACCUMULATE?"

CONNect

Command
:DISPlay:CONNect {{ON|1}|{OFF|0}}
The CONNect command sets the Connect Dots mode. When ON, each
displayed sample dot will be connected to the adjacent dot by a straight line.
The waveform is easier to see in this mode. When OFF, only the sampling
points will be displayed.
Example
OUTPUT XXX;":DISPLAY:CONNECT ON"