Record; Source - HP 16533A Programmer's Manual

1-gsa/s and 2-gsa/s digitizing oscilloscope
Hide thumbs Also See for 16533A:
Table of Contents

Advertisement

WAVeform Subsystem

RECord

RECord
Command
:WAVeform:RECord {FULL|WINDow}
The RECord command specifies the data you want to receive over the bus.
The choices are FULL or WINdow. When FULL is chosen, the entire 32768
point record of the specified channel is transmitted over the bus. In WINdow
mode, only the data displayed on screen will be returned.
Example
OUTPUT XXX;":WAV:SOUR CHAN1;REC FULL"
Query
:WAVeform:RECord?
The RECord query returns the present mode chosen.
Returned Format
[:WAVeform:RECord] {FULL|WINDow}<NL>
Example
OUTPUT XXX;":WAVEFORM:RECORD?"

SOURce

Command
:WAVeform:SOURce CHANnel<N>
The SOURce command specifies the channel that is to be used for all
subsequent waveform commands.
1 through the number of channels in the oscilloscope connected as one unit
<N>
(maximum of 8 channels), otherwise the maximum number of channels is 2
Example
OUTPUT XXX;":WAVEFORM:SOURCE CHANNEL1"
Query
:WAVeform:SOURce?
10–12