MARKer Subsystem
AVOLt
AVOLt
Command
:MARKer:AVOLt CHANnel<N>,<level>
The AVOLt command moves the A marker to the specified voltage on the
indicated channel.
1 through the number of channels in the oscilloscope connected as one unit
<N>
(maximum of 8 channels), otherwise the maximum number of channels is 2
the desired marker voltage level, ranging from ±(2 x maximum offset)
<level>
Example
OUTPUT XXX;":MARKER:AVOLT CHANNEL1,2.75"
Query
:MARKer:AVOLt?
The AVOLt query returns the current voltage and channel selection for the A
marker.
Returned Format
[:MARKer:AVOLt]CHANnel<N>,<level><NL>
Example
OUTPUT XXX;":MARKER:AVOLT?"
6–6