The MSTats query returns the current setting.
Returned Format
[:MARKer:MSTats]{1|0}<NL>
Example
OUTPUT XXX;":MARKER:MSTATS?"

OAUTo

Command
:MARKer:OAUTo{ MANual|CHANnel<N>,<type>,<level>,
<slope>,<occurrence>}
The OAUTo command specifies the automatic placement specification for
the O marker. The first parameter specifies if automarker placement is to be
in the manual mode or on a specified channel. If a channel is specified, four
other parameters must be included in the command syntax. The four
parameters are: marker type, level, the slope, and the occurrence count.
1 through the number of channels in the oscilloscope connected as one unit
<N>
(maximum of 8 channels), otherwise the maximum number of channels is 2
ABSolute or PERCent
<type>
percentage of waveform voltage level, ranging from 10 to 90 of the Vtop to
<level>
Vbase voltage or a voltage level
POSitive or NEGative
<slope>
integer from 1 to 100
<occurrence>
Example
OUTPUT XXX;":MARKER:OAUTO CHANNEL1,PERCent,50,POSITIVE,5"