Sensor Setup - ABB Navigator 550 Operating Instructions Manual

Hydrazine analyzer, measurement made easy
Hide thumbs Also See for Navigator 550:
Table of Contents

Advertisement

Navigator 500
Hydrazine analyzer

7.1.2 Sensor Setup

Menu
Sensor Setup
Select
Exit
Menu
Hydrazine 1 – 4
Tag
Units
Range High
Range Low
Filter Type
Filter Time
Solution Detection
Flow Meter
Man. Valve Control
High Standard Valve Enables the High Standard Valve for manual control.
Reset to Defaults
OI/AHM550–EN Rev. A
Used to set the wet-section tag, measurement units, operational range and filter type, enable /
disable the Flow Meter (optional) and energize the solenoid valve (via the Man.l Valve Control
menu).
Comment
The identity of the hydrazine wet-section being set up.
Sets the (colored) 16-character alphanumeric wet-section tag
displayed on Operator Pages.
Selects the measurement unit:
ppb / µg/l or µg/kg
Sets the span value in Chart and Bargraph views.
Sets the zero value in Chart and Bargraph views.
Selects the filter type:
Off / Min / Max / Average / Rolling Point
Sets the filter (input) time (5 to 100 seconds).
Enables / disables the solution detection function.
Generates a No Solution diagnostic if solution is not detected when
Enabled – see page 64.
Enables / Disables the flow sensor function (active only if the flow
sensor option is fitted).
The solenoid valve can be energized manually. The current
concentration value is shown for reference. When the Man. Valve
Control parameter is exited, the valve returns to the closed state.
Note. Press the
key to open the valve manually, press the
to close the valve manually.
Resets all parameters to their default values.
Default
TAG 1
ppb
1000
0
Off
5 seconds
Enabled
Enabled
key
Disabled
41

Advertisement

Table of Contents
loading

This manual is also suitable for:

Navigator 500 seriesNavigator 540

Table of Contents