Recalibrating Sensors; Cleaning The Modular Flue Gas Probe; Replacing Probe Preliminary Filter; Replacing Thermocouple - TESTO 340 Instruction Manual

Flue gas analyser
Hide thumbs Also See for 340:
Table of Contents

Advertisement

I.4
Recalibrating sensors
See Sensor settings, p. 31.
I.5
Cleaning the modular flue gas probe
Probe shafts with preliminary filter:
Blow compressed air through the preliminary filter. For thorough cleaning, use an
ultrasonic bath or a cleaner for dentures. Screw the preliminary filter back on to
the probe shaft after cleaning.
3 Fit a new probe module on the handle and engage it in place.
I.6
Replacing probe preliminary filter
The preliminary filter in probe modules fitted with a preliminary filter can be replaced.
Unscrew the preliminary filter from the probe shaft and screw on a new filter.
I.7 Replacing thermocouple
Detach the flue gas probe from the measuring inst-
rument before cleaning.
1 Release the probe catch by pressing the key on
the probe handle and remove the probe module.
Probe shafts with preliminary filter:
Unscrew the preliminary filter.
2 Blow compressed air through the flue ducts of the
probe module and probe handle (see illustration).
Do not use a brush!
1 Release the probe catch by pressing the key on
the probe handle and remove the probe module.
2 Detach the plug-in head of the thermocouple
from its mounting using a screwdriver and pull the
thermo couple from the probe shaft.
3 Lead a new thermocouple into the probe shaft until
the plug-in head engages.
4 Fit probe module on the handle and engage it in
place.
49