Sensor Settings; F.2 Sensor Settings - TESTO 340 Instruction Manual

Flue gas analyser
Hide thumbs Also See for 340:
Table of Contents

Advertisement

F.2 Sensor settings

It is possible to set an NO
(dilution/disconnect). The actual calibration data and the status of the sensors can be
displayed. Recalibration can be carried out.
Calling up the function:

Sensor settings

Ý
Setting the NO2 addition (as long as no NO2 sensor is plugged in):
1
.
NO2 addition
Option:
Reset N02 addition to default value:
2
Ý Set the value Ý
Change
Setting sensor protection:
To extend the measuring range and protect the sensors against overloads, you can set
thresholds which, when exceeded, activate sensor protection. Thresholds for a variety
of parameters can be set, depending on the sensors that are connected.
1
Sensor protection
2 Select the parameter.
Option:
Reset selected parameter to default value:
3
Ý Set the values Ý
Change
4 Repeat steps 2 and 3 for the other parameters accordingly.
Saving settings:
Measurement CO (H2-compensated) sensor:
In order to protect the sensor and for a longer sensor life, we recommend that in
measurements with unexpectedly high CO concentrations (more than 1,000ppm),
the CO sensor is installed in slot 2, and that the threshold of the CO sensor protecti-
on is set to 1,000ppm. From a CO concnetration of 1,000ppm, dilution with a factor
of 5 is automatically activated.
This setting can also be made if H2 concentrations of more than 1,000ppm are to be
expected.
Display ppm/hour counter (active only when sensors with exchangeable filters are
used):
For those sensors which have an exchangeable chemical filter for neutralizing cross-ga-
ses, a ppm/hour counter is available.
addition and thresholds for activating sensor protection
2
OK
.
Ý
.
OK
OK
.
Ý
.
OK
OK
OK Save input
Ý
Deflt
.
Deflt
.
.
33

Advertisement

Table of Contents
loading

Table of Contents