CHEMICAL ANALYSIS INFORMATION

Reagent Blank Correction

Standard Adjust (Adjusting the Standard Curve)

The Reagent Blank Correction subtracts the color absorbed when running
the test with deionized water instead of sample. The blank value is
subtracted from every result to correct for any background color due to
reagents.
When using the Reagent Blank Correction feature, the blank correction
should be entered before the Standard Adjust feature is used.
To enter a programmed correction for the reagent blank:
1. Run the test using deionized water with each new lot of reagents.
2. Press
to obtain the blank value.
READ
3. Press
, scroll to
SETUP
show
.
BLANK?
4. Enter the blank value just read from the instrument.
5. Press
to accept the value as the blank to be subtracted from
ENTER
each reading.
6. The display will show 0.00 mg/L (resolution and units vary) and the
sample cell icon will be displayed, indicating that the reagent blank
feature is enabled and the blank value will be subtracted from each
reading. Repeat the reagent blank adjust for each new lot of reagents.
Note: After entering a reagent blank adjust, the display may flash "limit" when zeroing if
the sample used for zeroing has a lower absorbance value than the reagent blank.
To disable the Reagent Blank adjust feature, press
and press
BLANK
ENTER
displayed without subtracting the blank. The sample cell icon will no
longer appear in the display.
Do not use the Reagent Blank Adjust feature if the procedure uses a
reagent blank for zeroing.
The colorimeter has Hach Programs permanently installed in memory. A
program usually includes a pre-programmed calibration curve. Each curve
is the result of an extensive calibration performed under ideal conditions
and is normally adequate for most testing. Deviations from the curve can
occur from using compromised testing reagents, defective sample cells,
incorrect test procedure, incorrect technique, or other correctable causes.
Interfering substances or other causes may be beyond the analyst's control.
, continued