Nitrate, Low Range, Cadmium Reduction Method - Hach DR/850 Procedures Manual

Colorimeter
Hide thumbs Also See for DR/850:
Table of Contents

Advertisement

NITRATE, Low Range
Cadmium Reduction Method
1.
Enter the stored
program number for low
range nitrate nitrogen
-
(NO
-N).
3
Press:
PRGM
The display will show:
PRGM ?
Note: For most accurate
results, perform a Reagent
Blank Correction using
deionized water (see
Section 1).
* Seawater requires a manual calibration; see Interferences.
(0 to 0.50 mg/L NO
2.
Press:
55 ENTER
The display will show
mg/L, NO3-N and the
ZERO icon.
Note: For alternate forms
(NO
), press the
3
CONC key.
-
-N)
3
For water, wastewater and seawater
3.
Fill a 25-mL
graduated mixing
cylinder to the 15-mL
mark with sample.
Note: Adjust the pH of
stored samples before
analysis.
243
Method 8192
4.
Add the contents of
one NitraVer 6 Nitrate
Reagent Powder Pillow
to the cylinder. Stopper.
Note: It is necessary to
remove all the powder from
the foil pillow. Tap the
pillow until no more
powder pours out. Be sure
to remove powder from the
corners of the pillow.
*