Hach DR/850 Procedures Manual page 275

Colorimeter
Hide thumbs Also See for DR/850:
Table of Contents

Advertisement

NITROGEN, AMMONIA, High Range, Test 'N Tube,
Summary of Method
Safety
Pollution Prevention And Waste Management
REQUIRED REAGENTS
AmVer
Reagent Set for Nitrogen, Ammonia, High Range, TNT (25 tests) ..............26069-45
Includes: (1) 23952-66, (1) 23954-66, (1) 272-42,
Description
AmVer
HR Reagent Test 'N Tube
Ammonia Salicylate Reagent Powder Pillows....... 2 pillows.................50/pkg ..........23952-66
Ammonia Cyanurate Reagent Powder Pillows ...... 2 pillows.................50/pkg ..........23954-66
REQUIRED APPARATUS
COD/TNT Adapter...........................................................1 .................each.................48464-00
®
Pipet, TenSette
, 0-1 mL .................................................1 .................each.................19700-01
Pipet Tips for 19700-01................................................ varies ......... 50/pkg.................21856-96
Test Tube Rack ............................................................... 1-3 ...............each.................18641-00
Funnel, micro (for reagent addition) ................................1 .................each.................25843-35
* Not available separately.
Ammonia compounds combine with chlorine to form
monochloramine. Monochloramine reacts with salicylate to form 5-
aminosalicylate. The 5-aminosalicylate is oxidized in the presence of a
sodium nitroprusside catalyst to form a blue-colored compound. The
blue color is masked by the yellow color from the excess reagent
present to give a green-colored solution.
Good safety habits and laboratory techniques should be used
throughout the procedure. Consult the Material Safety Data
Sheets for information specific to the reagents used. For
additional information, refer to Section 3.
The ammonia salicylate reagent contains sodium
nitroferricyanide. Cyanide solutions are regulated as hazardous
wastes by the Federal RCRA. Collect cyanide solutions for
disposal as reactive (D001) waste. Be sure cyanide solutions are
stored in a caustic solution with pH >11 to prevent release of
hydrogen cyanide gas. See Section 3 for further information in
proper disposal of these materials.
Quantity Required
Vials ............ 2 vials ...................50/pkg .......................*
275
*
(50) AmVer HR Vials
Per Test
continued
Unit
Cat. No.