Hach DR/850 Procedures Manual page 414

Colorimeter
Hide thumbs Also See for DR/850:
Table of Contents

Advertisement

SULFIDE,
continued
9.
Press:
ZERO
The cursor will move to
the right, then the
display will show:
0.00 mg/L S
Sampling and Storage
Accuracy Check
Method Performance
10.
After the timer
beeps, place the
prepared sample into the
cell holder. Tightly
cover the sample cell
with the instrument cap.
Collect samples in clean plastic or glass bottles. Fill completely
and cap tightly. Avoid excessive agitation or prolonged exposure
to air. Analyze samples immediately.
Sulfide standards are unstable and must be user prepared. See
Sandard Methods, 4500S
directions.
Precision
In a single laboratory, using standard solutions of 0.73 mg/L
sulfide and two representative lots of reagent with the instrument,
a single operator obtained a standard deviation of ± 0.02 mg/L
sulfide.
Estimated Detection Limit (EDL)
The EDL for program 93 is 0.01 mg/L S
on derivation and use of Hach's estimated detection limit, see
Section 1.
11.
Press:
READ
The cursor will move to
the right, then the result
in mg/L sulfide will be
displayed.
Note: Some sulfide loss
may occur if dilution is
necessary.
Note: Standard Adjust may
be performed using a
prepared standard (see
Standard Adjust in
Section 1).
for preparation and standardization
414
2-
. For more information