Hach DR/850 Procedures Manual page 380

Colorimeter
Hide thumbs Also See for DR/850:
Table of Contents

Advertisement

PHOSPHORUS, TOTAL,
9.
Carefully remove the
vial from the reactor.
Place it in a test tube
rack and allow to cool to
room temperature.
Note: Vials will be hot.
13.
Tightly cover the
vial with the instrument
cap.
continued
10.
Use a TenSette
Pipet to add 2.0 mL of
1.54 N sodium
hydroxide to the vial.
Cap and mix.
14.
Press:
ZERO
The cursor will move to
the right, then the
display will show:
0.00 mg/L PO4
Note: For multiple
samples, zero only on the
first sample. Read the
remaining samples after
adding the PhosVer 3
reagent.
11.
Clean the outside of
the vial with a towel.
Note: Wiping with a damp
towel, followed by a dry
one, will remove
fingerprints or other marks.
15.
Remove the cap
from the vial. Using a
funnel, add the contents
of one PhosVer 3
Phosphate Reagent
Powder Pillow to the
vial.
380
12.
Place the sample
vial in the adapter.
Push straight down on
the top of the vial until it
seats solidly into the
adapter.
Note: Do not move the vial
from side to side as this can
cause errors.
16.
Cap tightly and
shake for 10-15 seconds.
Note: The powder will not
completely dissolve.