Hach DR/850 Procedures Manual page 212

Colorimeter
Hide thumbs Also See for DR/850:
Table of Contents

Advertisement

IRON, TOTAL,
Interferences
Interfering Substance
Cadmium
3+
Chromium (
)
6+
Chromium (
)
Cobalt
Color or turbidity
Copper
Cyanide
Manganese
Mercury
Molybdenum
Nickel
Nitrite Ion
pH
Summary of Method
continued
Estimated Detection Limit
The estimated detection limit for program 39 is 0.04 mg/L Fe. For
more information on the estimated detection limit, see Section 1.
Greater than 4.0 mg/L
Greater than 0.25 mg/L
Greater than 1.2 mg/L
Greater than 0.05 mg/L
If the sample is turbid, add one 0.1-g scoop of RoVer Rust Remover to the blank in
Step 6 (Step 3 for AccuVac procedure). Swirl to mix.
Greater than 0.6 mg/L
Greater than 2.8 mg/L
Greater than 50.0 mg/L
Greater than 0.4 mg/L
Greater than 4.0 mg/L
Greater than 1.0 mg/L
Greater than 0.8 mg/L
A sample pH of < 3 or >4 after the addition of reagent may inhibit color formation,
cause the developed color to fade quickly or result in turbidity. Adjust the sample pH
to 3–5 before adding reagent using a pH meter or pH paper and adding (dropwise) an
appropriate amount of iron-free acid or base (i.e., 1.0 N Sulfuric Acid Standard
Solution or 1.0 N Sodium Hydroxide Standard Solution). Make a volume correction if
significant volumes of acid or base are used.
The TPTZ Iron Reagent forms a deep blue-purple color with
ferrous iron. The indicator is combined with a reducing agent
which converts precipitated or suspended iron, such as rust, to the
ferrous state. The amount of ferric iron present can be
determined as the difference between the results of a ferrous iron
test and the concentration of total iron.
Interference Levels and Treatments
212