FLUORIDE,
continued
9.
Place the prepared
sample into the cell
holder. Tightly cover the
sample cell with the
instrument cap.
Using AccuVac Ampuls
1.
Enter the stored
program number for
-
fluoride (F
)- AccuVac
Ampuls.
Press:
PRGM
The display will show:
PRGM ?
10.
Press:
READ
The cursor will move to
the right, then the result
in mg/L fluoride will be
displayed.
Note: Use of the Standard
Adjust feature with each
new lot of reagent is highly
recommended. See
Accuracy Check following
these steps.
2.
Press:
28 ENTER
The display will show
mg/L, F and the
ZERO icon.
3.
Collect at least 40
mL of sample in a
50-mL beaker. Pour at
least 40 mL of
deionized water into a
second beaker.
170
4.
Fill a SPADNS
Fluoride Reagent
AccuVac Ampul with
sample by breaking the
tip on the bottom of the
beaker. Fill a second
AccuVac Ampul with
deionized water
(the blank).
Note: Keep the tip
immersed while the ampule
fills completely.