Nitrogen, Ammonia, Low Range, Test 'N Tube Salicylate Method - Hach DR/850 Procedures Manual

Colorimeter
Hide thumbs Also See for DR/850:
Table of Contents

Advertisement

NITROGEN, AMMONIA, Low Range, Test 'N Tube
Salicylate Method
1.
Enter the stored
program number for low
range nitrogen, ammonia
Test 'N Tube.
Press:
PRGM
The display will show:
PRGM ?
5.
Using a funnel, add
the contents of one
Ammonia Salicylate
Reagent Powder Pillow
for 5 mL sample to each
vial.
* Adapted from Clin. Chim. Acta, 14 403 (1966).
*
2.
Press:
66 ENTER
The display will show
mg/L, NH3-N and the
ZERO icon.
Note: For alternate forms
(NH
), press the
3
CONC key.
6.
Using a funnel, add
the contents of one
Ammonia Cyanurate
Reagent Powder Pillow
for 5 mL sample to each
vial.
For water, wastewater, and seawater
3.
Insert the COD/TNT
Adapter into the cell
holder by rotating the
adapter until it drops
into place. Then push
down to fully insert it.
Note: For increased
performance, a diffuser
band covers the light path
holes on the adapter. Do
not remove the diffuser
band.
7.
Cap the vials tightly
and shake thoroughly to
dissolve the powder.
Note: A green color
will develop if ammonia is
present.
265
Method 10023
(0 to 2.50 mg/L NH
4.
Remove the caps
from 2 AmVer Diluent
Reagent vials. Add 2 mL
of sample to one vial
(the sample). Add 2 mL
of deionized water to the
other vial (the blank).
Note: Adjust the pH of
stored samples before
analysis. See Interferences
on page 267.
8.
Press:
TIMER ENTER
A 20-minute reaction
period will begin.
-N)
3