Chlorine, Total, Ultra-High Range, Dpd Method - Hach DR/850 Procedures Manual

Colorimeter
Hide thumbs Also See for DR/850:
Table of Contents

Advertisement

CHLORINE, TOTAL, Ultra-High Range
DPD Method
Note: This product has not been evaluated to test for chlorine and chloramines in medical
applications in the United States.
1.
Enter the user
program number for
Chlorine, UHR.
Press:
PRGM
The display will show:
PRGM?
Note: If the chlorine is
typically less than 2.0 mg/
L, use method 8167,
program number 9.
* Procedure is equivalent to USEPA method 330.5 for wastewater and Standard Method 4500-C1-G for drinking
water.
USEPA accepted for reporting water and wastewater analyses
For testing higher levels of total chlorine (free and combined)
industrial process waters, or treated wastewater
2.
Press:
12 ENTER
The display will show
mg/L Cl
2
then: ZERO
(0.0–10.0 mg/L Cl
in drinking water, cooling water,
10
mL
5
mL
3.
Fill the 10-mL/1-cm
cell to the 5-mL line
with sample.
109
) Method 10070
2
Tab
Slot
4.
Place the sample cell
into the instrument.
Cover the sample cell
tightly with the
instrument cap.
Note: Place the cell into
the cell holder as
illustrated. The sample cell
tab should be at the 6
o'clock position and
completely seated in the
cell holder slot.
*