NITROGEN, AMMONIA, High Range, Test 'N Tube,
Interferences
The following ions may interfere when present in concentrations
exceeding those listed below.
In some lab environments, airborne cross contamination of the
blank is possible. Complete preparation of the blank before
opening or handling any samples or standards to avoid transfer of
ammonia. If sample or standard containers have already been
open, move to a separate area of the lab to prepare the blank.
Substance
Acidic or basic
Adjust to approximately pH 7. Use 1 N Sodium Hydroxide
samples
Standard Solution for acidic samples and 1 N Hydrochloric Acid
Standard Solution for basic samples.
Calcium
50,000 mg/L as CaCO
Glycine, hydrazine
Will cause intensified colors in the prepared sample.
Magnesium
300,000 mg/L as CaCO
Iron
Eliminate iron interference as follows:
1. Determine the amount of iron present in the sample using one
2. Add the same iron concentration to the deionized water in step
3. The interference will then be successfully blanked out.
Nitrite
600 mg/L as NO
Nitrate
5,000 mg/L as NO
Orthophosphate
5,000 mg/L as PO
Sulfate
6,000 mg/L as SO
Sulfide
Sulfide will intensify the color. Eliminate sulfide interference as
follows:
1. Measure about 350 mL of sample in a 500 mL Erlenmeyer
2. Add the contents of one Sulfide Inhibitor Reagent Powder
3. Filter the sample through folded filter paper. Use the filtered
Turbidity and color
Give erroneous high values. Samples with severe interferences
require distillation. Hach recommends the distillation procedure
using the Hach General Purpose Distillation Set.
Concentration and Suggested Treatments
3
3
of the total iron procedures.
4.
-
-N
2
-
-N
3
3-
-P
4
2-
4
flask.
Pillow. Swirl to mix.
solution in step 4.
274