PHOSPHORUS, REACTIVE,
Using AccuVac Ampuls
1.
Enter the stored
program number for
reactive phosphorus-
ascorbic acid method.
Press:
PRGM
The display will show:
PRGM ?
Note: For most accurate
results, perform a Reagent
Blank Correction using
deionized water (see
Section 1).
continued
2.
Press:
79 ENTER
The display will show
mg/L, PO4 and the
ZERO icon.
Note: For alternate forms
(P, P
O
), press the CONC
2
5
key.
3.
Fill a sample cell
(the blank) with at least
10 mL of sample.
Collect at least 40 mL
of sample in a 50-mL
beaker.
Note: For samples with
extreme pH, see
Interferences.
Note: Clean glassware
with 1:1 HCl. Rinse again
with deionized water. Do
not use detergent
containing phosphates to
clean glassware.
349
4.
Fill a PhosVer 3
Phosphate AccuVac
Ampul with sample.
Note: Keep the tip
immersed while the ampul
fills completely.