Technical data

9.1

Fuse values

Attention
Values
IEC
CSA/UL
CSA/UL
9.2

Electric cable size

Attention
IEC
CSA/UL
CSA/UL
9.3

Device settings

Regulating and safety devices
The regulating and safety devices are factory-adjusted to obtain optimum performance of the dryer. Do not
alter the setting of any of the devices.
80
The indicated fuse value is the maximum value. The cable size used may specify fuses of
a lower value.
Frequency
Voltage
50 Hz
230 V
60 Hz
115 V
60 Hz
230 V
Local regulations remain applicable if they are stricter than the values proposed below.
The voltage drop must not exceed 5% of the nominal voltage. It may be necessary to use
cables of a larger size than those stated to comply with this requirement.
Frequency
50 Hz
60 Hz
60 Hz
Fuse specification
gL/gG
CSA HRC/UL RK5
CSA HRC/UL RK5
Voltage
230 V
115 V
230 V
Instruction book
Maximum fuse value
2 A
2 A
2 A
Minimum cable size
3 x 1.5 mm²
3 x AWG14
3 x AWG14
2920 7099 70