10
Technical data
10.1 Fuse values
Attention
Values
IEC
CSA/UL
CSA/UL
10.2 Electric cable size
Attention
IEC
CSA/UL
CSA/UL
10.3 Device settings
Regulating and safety devices
The regulating and safety devices are factory-adjusted to obtain optimum performance
of the dryer. Do not alter the setting of any of the devices.
80
The indicated fuse value is the maximum value. The cable size used may specify fuses of
a lower value.
Frequency
Voltage
50 Hz
230 V
60 Hz
115 V
60 Hz
230 V
Local regulations remain applicable if they are stricter than the values proposed below.
The voltage drop must not exceed 5% of the nominal voltage. It may be necessary to use
cables of a larger size than those stated to comply with this requirement.
Frequency
50 Hz
60 Hz
60 Hz
Fuse specification
gL/gG
CSA HRC/UL RK5
CSA HRC/UL RK5
Voltage
Minimum cable size
230 V
3 x 1.5 mm²
115 V
3 x AWG14
230 V
3 x AWG14
Maximum fuse value
2 A
2 A
2 A