BeaconMedaes dMED 7 Instruction Book page 69

Dual medical air purifiers
Table of Contents

Advertisement

Instruction book
Stopping
To stop the air treatment unit, proceed as follows:
1. Close the external inlet valve between the air receiver and the dMED and the external outlet valve between
the dMED and the air net.
2. Let the unit operate for a period without consumption, to depressurize the vessels.
3. Stop the dMED (ES-MED controller).
2920 7099 70
If the air treatment unit is stopped for a longer period, keep the external inlet and outlet
valve closed to avoid moisture from entering the unit.
Under no circumstances must compressed air be allowed to flow through the unit when
the electrical power is switched off. This will result in terminal failure of the desiccant
material, causing regeneration will no longer be possible.