The spinning circle symbol will disappear, indicating that the ES-MED is not active (controlling the
compressors) any more. The compressors are now in Forced Local Mode even though their local/LAN switch
may be in LAN position. This allows the operator to do maintenance or troubleshooting on the central
controller while pressure is guaranteed by the local compressors.
2920 7099 70
31