Reference conditions and limitations

Reference conditions
Compressed air inlet pressure
Compressed air inlet pressure
Compressed air inlet temperature
Compressed air inlet temperature
Ambient temperature
Ambient temperature
Inlet relative air humidity
Pressure dew point
Pressure dew point
Limitations
Maximum compressed air inlet pressure
Maximum compressed air inlet pressure
Minimum compressed air inlet pressure
Minimum compressed air inlet pressure
Min.-max. ambient air temperature
Min.-max. ambient air temperature
Min.-max. compressed air inlet temperature
Min.-max. compressed air inlet temperature
Maximum concentration of contaminants:
O
concentration
2
CO
concentration
2
CO concentration (without catalyst (QDT))
CO concentration (with catalyst (QDT+))
SO
concentration
2
NO
concentration
x
Water vapour amount
Oil vapour amount
2920 7099 70
Units
bar(e)
See section
psig
See section
˚C
35
˚F
95
˚C
25
˚F
77
%
100
°C
-40
°F
-40
Units
bar(e)
See section
psig
See section
bar(e)
See section
psig
See section
°C
2 - 45
°F
35.6 - 113
°C
2 - 50
°F
33.8 - 122
%
20.4 < x < 21.4
ppm
700
ppm
0
ppm
50
ppm
5
ppm
5
%
100
3
3
mg/m
dMED data
dMED data
dMED data
dMED data
dMED data
dMED data
81