BeaconMedaes dMED 7 Instruction Book page 34

Dual medical air purifiers
Table of Contents

Advertisement

To restart after having stopped the ES-MED (see above), proceed to paragraph Starting to start the central
controller, putting it in charge of the compressors again.
Isolating a compressor controller
To carry out maintenance on a compressor without transmitting alarms, select the ES submenu from the main
menu (requires full access, see above). Browse to the compressor that needs to be isolated and press Enter.
In the following selection menu, navigate to <Mode> and select <Isolated>.
Important:
After maintenance, this process shall be followed to select the mode to <Integrated> again.
32
Instruction book
2920 7099 70

Hide quick links: