BeaconMedaes dMED 7 Instruction Book page 81

Dual medical air purifiers
Table of Contents

Advertisement

Instruction book
Condition
Pressure sensors: error signal
(no measurement)
Pressure sensors: too high or
too low indication
Pressure regulator: wrong
outlet pressure
Dew point sensor
malfunctioning
Dew point sensor: no
measurement (error signal)
Dew point sensor: incorrect
measurement
Faults and remedies, ES-MED
Condition
Doesn't read in pressures
Doesn't read in (correct)
compressor status
Doesn't produce run queue
Doesn't operate the
compressors
Gives out <Run> or <Not run>
command too slowly
Doen't issue alarm(s) when it
should
Doesn't issue service warnings
when it should
Keeps exhaust (or inlet) valves
closed (or opened)
2920 7099 70
Fault
Electric connection
interrupted
Mechanical damage (e.g.
too high mounting torque)
Connection plugs not
connected
Crack due to incorrect
mounting
Internal defect
No supply voltage
Condense on connection
plugs
Faulty supply signal
Clogging or contamination
Sensor internally defect
Deviation due to ageing
Deviation due to incorrect
production
Wrong setting
No direct feedback of dew point Replace defective parts
No direct feedback of dew point
Dew point alarm generated
too early or too late
Fault
Software fault
Software fault
Software fault
Software fault
Communication fault
Software fault
Software fault
Connection fault between
ES-MED and hospital
control room display
Software fault
Software fault
Communication fault
Remedy
Check connections
Check sensors
Check sensors
Replace defective parts
Check gauge of pressure regulator
Adjust pressure
Check connections
Check sensors
Check sensors
Replace defective parts
Remedy
Reprogram the ES-MED
Reprogram the ES-MED
Reprogram the ES-MED
Reprogram the ES-MED
Check wiring
Reprogram the ES-MED
Reprogram the ES-MED
Check wiring
Reprogram the ES-MED
Reprogram the ES-MED
Check wiring
Check solenoid