3
Using the M8040A High-Performance BERT

Updating Software Components

Contacting Keysight Service and Support

Quick Troubleshooting

66
Updated versions of the M8070A and module specific software
components are available on the Keysight website.
These software components are available as .EXE files. To download a
software upgrade:
1
Go to
http://www.keysight.com/find/m8070a
2
Click the Technical Support tab.
3
Click Drivers and Software.
4
Type the model number of the instrument module for which software
update is needed and click Find. Model number is located on the front
panel of the module.
5
Click the Driver and Software link on the module page.
6
Download the required software update from the list of available
updates.
To locate a sales or a service office near you, go to
www.keysight.com/find/contactus
This section lists the basic troubleshooting steps that you might find useful
while setting up the M8040A.
The chassis does not power up
f the chassis or a module does not appear to power up, check the
following:
The circuit breakers at the rear of the chassis are set to the right, which
is the ON position.
The AC power cords are connected to a working power source.
The electrical circuits are not overloaded. Check the combined power
requirements of all equipment on the same circuit.
There are no empty slots in the chassis. Leaving slots empty can
overheat the inserted modules, causing them to shut down.