Summary of Contents for Keysight Technologies M8000 Series
Page 1
Keysight M8000 Series of BER Test Solutions J-BERT M8020A High-Performance BERT M8030A Multi-Channel BERT M8040A High-Performance BERT Installation Guide...
Page 2
ALL WARRANTIES, EITHER EXPRESS OR its standard commercial license, which is met. IMPLIED WITH REGARD TO THIS MANUAL embodied in its End User License Agree- AND ANY INFORMATION CONTAINED Keysight M8000 Series of BER Test Solutions Installation Guide...
Failure to comply with these precautions or with specific warnings or operating instructions in the product manuals violates safety standards of design, manufacture, and intended use of the instrument. Keysight Technologies assumes no liability for the customer's failure to comply with these requirements. Product manuals are provided with your instrument on CD-ROM and/or in printed form.
Page 4
(Australian Spectrum Management Agency). This indicates compliance with all Australian EMC regulatory information. Indicates that the product was tested and has met the certification requirements for electrical, plumbing and/or mechanical products. Keysight M8000 Series of BER Test Solutions Installation Guide...
Page 5
ISM GRP-1A classification according to the international EMC standard. ICES/NMB-001 compliance marking to the Canadian EMC standard. This symbol on all primary and secondary packaging indicates compliance to China standard GB 18455-2001. Keysight M8000 Series of BER Test Solutions Installation Guide...
“Monitoring and Control instrumentation” product. Do not dispose in domestic household waste. To return unwanted products, contact your local Keysight office, or see http://about.keysight.com/en/companyinfo/environment/takeback.shtml for more information. Keysight M8000 Series of BER Test Solutions Installation Guide...
“bundled” systems such as the M8020A-BU1, M8020A-BU2, M8030A-BU1, M8030A-BU2, M8040A-BU1, M8040A-BU2 or M8040A-BU3. For these systems, refer to the Keysight M8000 Series BER Test Solutions Getting Started Guide and Keysight M8040A Getting Started Guide. After performing the procedures in this document, you are directed to procedures in the Keysight M8000 Series BER Test Solutions Getting Started Guide and Keysight M8040A Getting Started Guide to complete the installation.
M8062A 32Gb/s Front-end for J-BERT M8020A High-Performance BERT Single M8020A LAN Network Configuration Hardware Configurations for M8030A Hardware Configurations for M8040A M8045A Pattern Generator Module M8046A Analyzer Module M8057A/B Remote Head M8040A LAN Network Configuration Keysight M8000 Series of BER Test Solutions Installation Guide...
Page 10
To Connect the M8057A/B Remote Head To remove an M8040A module Setting up the M8040A Benchtop Configuration Rack Mounted Configuration Installing the AXIe Embedded Controller Module Completing the Installation and Setup Process Keysight M8000 Series of BER Test Solutions Installation Guide...
Page 11
3 Controlling Multiple Instruments Controlling Multiple Instruments by Multiple Instances of M8070B System Software Remote Control of Multiple M8070B Software Instances Command Line Options Creating M8070B Software Shortcut on the Desktop Index Keysight M8000 Series of BER Test Solutions Installation Guide...
Page 13
Keysight M8000 Series of BER Test Solutions Installation Guide System Requirements / 14 Hardware and Software Requirements / 15 Hardware Configurations for M8020A / 20 Hardware Configurations for M8030A / 22 Hardware Configurations for M8040A This chapter provides information about possible configurations for...
M8040A systems. In case of M8020A, PCIe connectivity between the M9505A AXIe Chassis NOTE and an external desktop PC controller is recommended when full channel plus large patterns need to be downloaded. Keysight M8000 Series of BER Test Solutions Installation Guide...
M8041A module. The M8041A occupies three slots. A maximum of two M8020A modules can be installed in a 5-slot chassis. Figure 1 M8020A configuration for 16 Gb/s BERT with 1 to 2 channel Keysight M8000 Series of BER Test Solutions Installation Guide...
M8041A SYNC OUT to the M8051A SYNC IN to synchronize the two modules to a common system clock. Figure 2 M8020A configuration for 16 Gb/s BERT for up to 4 channels Keysight M8000 Series of BER Test Solutions Installation Guide...
5-slot M9505A AXIe Chassis, an M8041A module, and an M8061A module. The M8041A occupies three slots and the M8061A occupies two slots. Figure 3 M8020A configuration for 32 Gb/s BERT (external demultiplexer is recom- mended) Keysight M8000 Series of BER Test Solutions Installation Guide...
A typical configuration for an M8020A 32 Gb/s full BERT consists of the 5-slot M9505A AXIe Chassis, an M8041A module, and an M8062A module. The M8041A occupies three slots and the M8062A occupies two slots. Figure 4 M8020A configuration for 32 Gb/s BERT Keysight M8000 Series of BER Test Solutions Installation Guide...
Host PC with M8070B Client PC USB Connection (remote desktop or VNC) Client PC Host PC can also be M9536A or M9537A AXIe Embedded Controller Figure 5 Single M8020A LAN network configuration Keysight M8000 Series of BER Test Solutions Installation Guide...
M8051A module M9521A AXIe system module, always included in M8030A-BU1 or M8030A-BU2, mandatory # 8-9 M8051A module # 10-11 M8051A module # 12-13 M8051A module # 14 M8192A multi-channel synchronization module, mandatory Keysight M8000 Series of BER Test Solutions Installation Guide...
Page 21
System Requirements Figure 6 on page -21 shows an example of modules arrangement in the M9514A AXIe 14-slot chassis. Figure 6 Example of M8030A module arrangement Keysight M8000 Series of BER Test Solutions Installation Guide...
The M8045A module must be installed in slots 1 through 3 of the NOTE M9505A AXIe chassis for proper local bus communication unless the M9537A AXIe Embedded Controller is installed (must be in slot 1). Keysight M8000 Series of BER Test Solutions Installation Guide...
The following figure illustrates an M8057A/B remote head connected with one channel of M8045A module. M8057A/B Remote Head Figure 9 M8045A, M8046A and M8057A/B configuration Keysight M8000 Series of BER Test Solutions Installation Guide...
Client PC Host PC with M8070B Client PC USB Connection (remote desktop or VNC) Host PC can also be M9537A AXIe Embedded Controller Client PC Figure 10 M8040A LAN network configuration Keysight M8000 Series of BER Test Solutions Installation Guide...
Keysight M8000 Series of BER Test Solutions Installation Guide Installing Modules / 26 AXIe Instrument Modules / 27 Installing M8020A Module(s) / 39 Setting up the M8020A / 43 Setting up the M8030A / 55 Installing M8040A Module(s) / 63...
The drawing below illustrates the AXIe module’s general layout, backplane connections and chassis fasteners, viewed from the top. Test connections are made at the module’s front panel. The front panel and backplane connectors will vary depending on the module. Keysight M8000 Series of BER Test Solutions Installation Guide...
• The enclosure surface of the instrument module may become hot during use. If you need to remove the module, first power down the AXIe chassis, allow the module to cool, and then pull the module out of the chassis. Keysight M8000 Series of BER Test Solutions Installation Guide...
Then gently pull the module out of the chassis holding the screws. Retaining Screw Filler Panel Module Removed Figure 12 Filler panel module removed Keysight M8000 Series of BER Test Solutions Installation Guide...
Page 29
This allows proper air flow and cooling, and provides EMI shielding for the chassis and installed components. Leaving slots empty can increase fan speed, raise ambient noise, overheat components, and can cause the module to shut down. Keysight M8000 Series of BER Test Solutions Installation Guide...
Then gently pull the module out of the chassis holding the screws. Retaining Screw Filler Panel Module Removed Figure 15 Filler panel module removed Keysight M8000 Series of BER Test Solutions Installation Guide...
Page 31
Tighten retaining screw Figure 16 Tighten retaining screws Locate the synchronization cable that was shipped with the M8051A module as shown in Figure 17 on page -31. Figure 17 Standard synchronization cable M8041-61601 Keysight M8000 Series of BER Test Solutions Installation Guide...
Page 32
Route the synchronization cable as shown in Figure 18 on page -32 using the two self-adhesive cable holders. Sync Cable Self-adhesive Holders Figure 18 Synchronization cable routing Keysight M8000 Series of BER Test Solutions Installation Guide...
Then gently pull the module out of the chassis holding the screws. Retaining Screw Filler Panel Module Removed Figure 20 Filler panel module removed Keysight M8000 Series of BER Test Solutions Installation Guide...
Page 34
Push the handles ends towards the edge of the chassis to tuck them away. Tighten the retaining screws on either end of the module to ensure the ground connection. Tighten retaining screw Figure 21 Tighten retaining screws Keysight M8000 Series of BER Test Solutions Installation Guide...
Then gently pull the module out of the chassis holding the screws. Retaining Screw Filler Panel Module Removed Figure 23 Filler panel module removed Keysight M8000 Series of BER Test Solutions Installation Guide...
Page 36
Push the handles ends towards the edge of the chassis to tuck them away. Tighten the retaining screws on either end of the module to ensure the ground connection. Tighten retaining screw Figure 24 Tighten retaining screws Keysight M8000 Series of BER Test Solutions Installation Guide...
This unseats the module from the chassis backplane. Once the module is unseated, use the module insertion/extraction handles by pulling directly outward to remove the module from the chassis. Keysight M8000 Series of BER Test Solutions Installation Guide...
Page 38
Installing Modules Do not remove the AXIe ESM, which is integral to the operation of the CAUTION chassis. An AXIe ESM that needs servicing should be removed by Keysight personnel only. Keysight M8000 Series of BER Test Solutions Installation Guide...
If you want to use the M8020A in a benchtop configuration, then retain the plastic bumpers and carry handle(s) for benchtop use. Bumpers Carry Handle Figure 27 Chassis bumpers and carry handles Keysight M8000 Series of BER Test Solutions Installation Guide...
Attach the rack mount brackets to the chassis and mount onto a rack. The rack mount brackets are available in the Keysight rack mount kit that you can order for the chassis. Refer to this kit for rack mounting instructions. Keysight M8000 Series of BER Test Solutions Installation Guide...
You do not need to manually install any operating system or drivers for this module. The Windows 7 (64 bit) operating system is pre-installed based on your choice and the module is ready to use as the host computer. Keysight M8000 Series of BER Test Solutions Installation Guide...
Page 42
NOTE the same time. If you plan to use the M9536A or M9537A AXIe Embedded Controller as the host computer, then do not connect an external host computer to the M8020A. Keysight M8000 Series of BER Test Solutions Installation Guide...
The enclosure surface of the instrument module may become hot during use. If you need to remove the module, first power down the AXIe chassis, allow the module to cool, and then pull the module out of the chassis. Keysight M8000 Series of BER Test Solutions Installation Guide...
Page 44
Hot Swap—AXIe does not explicitly support hot swap for instrument modules. Keysight recommends fully powering down the chassis before installing or removing modules. AXIe System Module—Shutdown of the controller operating system and power off the chassis before removing. Keysight M8000 Series of BER Test Solutions Installation Guide...
Page 45
Installing Modules Figure 30 M9514A AXIe 14-slot chassis at a glance Detailed information, consisting of user guide and other documents for the M9514A AXIe chassis is available at www.keysight.com/find/M9514A. Keysight M8000 Series of BER Test Solutions Installation Guide...
Keysight I/O libraries suite. Do not use the ASM LAN port as the remote control port. Use the LAN port of the NOTE controller (on M9536A or M9537A or external PC). Keysight M8000 Series of BER Test Solutions Installation Guide...
Page 47
The following figure displays the M9536A Embedded Controller installed in slot 1 of the M9514A AXIe chassis. Figure 31 M9536A embedded controller installed in M9514A AXIe chassis Keysight M8000 Series of BER Test Solutions Installation Guide...
Page 48
NOTE the same time. If you plan to use the M9536A or M9537A AXIe Embedded Controller as the host computer, then do not connect an external host computer to the M8030A. Keysight M8000 Series of BER Test Solutions Installation Guide...
Locate the (top and bottom) guide rails for each slot. The example below shows the guide rails in the chassis with all slots empty; typically one or more will be covered. Figure 32 Guide rails in the chassis with all slots empty Keysight M8000 Series of BER Test Solutions Installation Guide...
Page 50
Figure 33 Top view of left handle If either handle is misaligned, you will not be able to properly install the module. Keysight M8000 Series of BER Test Solutions Installation Guide...
Page 51
This allows proper air flow and cooling, and provides EMI shielding for the chassis and installed components. Leaving slots empty can increase fan speed, raise ambient noise, overheat components, and can cause the module to shut down. Keysight M8000 Series of BER Test Solutions Installation Guide...
M8192A SYNC OUT to the individual SYNC IN ports of the M8041A and the four M8051A modules in order to synchronize the data channels of all modules to a common system clock. Keysight M8000 Series of BER Test Solutions Installation Guide...
Page 53
Installing Modules Figure 34 Cross-wired synchronization cable M8041-61614 Route the synchronization cable as shown in the following figure. Figure 35 Synchronization cable routing Keysight M8000 Series of BER Test Solutions Installation Guide...
Page 54
NOTE (e.g. M8051A, M8062A) to the test setup, make sure to connect the sync cable after completing the other connections and also remove the sync cable first while disconnecting the connections. Keysight M8000 Series of BER Test Solutions Installation Guide...
• The enclosure surface of the instrument module may become hot during use. If you need to remove the module, first power down the AXIe chassis, allow the module to cool, and then pull the module out of the chassis. Keysight M8000 Series of BER Test Solutions Installation Guide...
Then gently pull the module out of the chassis holding the screws. Retaining Screw Filler Panel Module Removed Figure 37 Filler panel module removed Keysight M8000 Series of BER Test Solutions Installation Guide...
Page 57
This allows proper air flow and cooling, and provides EMI shielding for the chassis and installed components. Leaving slots empty can increase fan speed, raise ambient noise, overheat components, and can cause the module to shut down. Keysight M8000 Series of BER Test Solutions Installation Guide...
Then gently pull the module out of the chassis holding the screws. Retaining Screw Filler Panel Module Removed Figure 40 Filler panel module removed Keysight M8000 Series of BER Test Solutions Installation Guide...
Page 59
Push the handle ends towards the edge of the chassis to tuck them away. Tighten the retaining screws on either end of the module to ensure the ground connection. Tighten retaining screw Figure 41 Tighten retaining screws Keysight M8000 Series of BER Test Solutions Installation Guide...
M8057A/B. Also, make sure NOT to remove the M8057A/B connections when it is powered on. However, if you wish to remove the M8057A/B connections, ensure that the instrument is powered off. Keysight M8000 Series of BER Test Solutions Installation Guide...
This unseats the module from the chassis backplane. Once the module is unseated, use the module insertion/extraction handles by pulling directly outward to remove the module from the chassis. Keysight M8000 Series of BER Test Solutions Installation Guide...
Page 62
Installing Modules Do not remove the AXIe ESM, which is integral to the operation of the CAUTION chassis. An AXIe ESM that needs servicing should be removed by Keysight personnel only. Keysight M8000 Series of BER Test Solutions Installation Guide...
Attach the rack mount brackets to the chassis and mount onto a rack. The rack mount brackets are available in the Keysight rack mount kit that you can order for the chassis. Refer to this kit for rack mounting instructions. Keysight M8000 Series of BER Test Solutions Installation Guide...
You do not need to manually install any operating system or drivers for this module. The Windows 7 (64 bit) operating system is pre-installed based on your choice and the module is ready to use as the host computer. Keysight M8000 Series of BER Test Solutions Installation Guide...
Page 65
You should not connect the M8040A to multiple operating host computers at NOTE the same time. If you plan to use the M9537A AXIe Embedded Controller as the host computer, then do not connect an external host computer to the M8040A. Keysight M8000 Series of BER Test Solutions Installation Guide...
Starting the M8070B software interface • Making a basic measurement Detailed information, consisting of User Guide, Programming Guide and other documents for the Keysight M8020A, M8030A and M8040A are available at: www.keysight.com/find/M8020A www.keysight.com/find/M8030A www.keysight.com/find/M8040A Keysight M8000 Series of BER Test Solutions Installation Guide...
Keysight M8000 Series of BER Test Solutions Installation Guide Controlling Multiple Instruments Controlling Multiple Instruments by Multiple Instances of M8070B System / 70 Software / 74 Remote Control of Multiple M8070B Software Instances / 75 Creating M8070B Software Shortcut on the Desktop This chapter describes how the M8070B system software allows a PC to control multiple instruments.
Page 68
Ref Clk Out port of the M8045A (frame 1) is connected to the Ref Clk In port of M8045A (fame Figure 47 PC connected to AXIe 5 slot frame Keysight M8000 Series of BER Test Solutions Installation Guide...
Page 69
Also, it is also possible to connect a PC to large AXIe frame (14 slots) as shown in the following figure: Figure 49 PC connected to AXIe 14 slot frame All these configurations are controlled by a single instance of a M8070B system software. Keysight M8000 Series of BER Test Solutions Installation Guide...
Following are some possible configurations: One PC connected to multiple instruments Figure 50 on page -70 shows one PC connected to multiple instruments via USB. Figure 50 PC connected to multiple instruments via USB Keysight M8000 Series of BER Test Solutions Installation Guide...
Page 71
-71 shows two instruments (shaded transparent rectangles) hosted by a M9514A and controlled by a single PC only. Figure 51 PC connected to AXIe frame 14 slot via PCIe Keysight M8000 Series of BER Test Solutions Installation Guide...
Page 72
PCIe connection to the PC while multiple M8070B instances are connected via USB connection to the PC. See Figure 52 on page -72. Figure 52 PC connected to multiple instruments via USB and PCIe Keysight M8000 Series of BER Test Solutions Installation Guide...
Page 73
If you restart the M8070B system software with the same command line arguments, it will address the same frame again. In other words as long as the set-up is not changed e.g. /chassisid 1 will address always the same frame. Keysight M8000 Series of BER Test Solutions Installation Guide...
Slot Number – defines a slot number of the first module to constitute an instrument /workspace Workspace Name – defines a name for a workspace in which the settings are stored Keysight M8000 Series of BER Test Solutions Installation Guide...
Rename the created shortcut to a meaningful name, right click and select in the context menu Properties. Choose the Shortcut tab and change the Target entry field as shown in the following example: "C:\Program Files\Keysight\M8070B\bin\KeysightM8070B.exe" /chassisid 1 /slotnumber 1 /workspace "InstrumentOne" Keysight M8000 Series of BER Test Solutions Installation Guide...
Page 76
Controlling Multiple Instruments Figure 56 Changing shortcut target Click Apply. You can now start the ‘configured’ M8070B system software by a simple double click on the desktop icon. Keysight M8000 Series of BER Test Solutions Installation Guide...