Recommended action
Replace the power module.
6.2.8
Database Storage Error
The device has failed to write to the database memory. Any data written during this time
may have been lost.
Recommended actions
1. Reset the device by removing and reinstalling the power module.
2. If logging dynamic data is not needed, the advisory can be safely ignored.
3. If the condition persists, replace the transmitter.
6.2.9
Invalid Configuration
The device has detected a configuration error based on a user change to the device.
Recommended actions
1. In Field Communicator or AMS Wireless Configurator, go to Service Tools → Active
2. Correct the parameter that has a configuration error.
3. Reset the device by removing and reinstalling the power module.
4. If the condition persists, replace the transmitter.
6.2.10
HI-HI Alarm
The primary variable has surpassed the user-defined limit.
WARNING
Always respond to all indications of dangerous concentrations of toxic gas as though they
are real until they are proven false.
Recommended actions
1. Verify the process variable is within user specified limits.
2. Reconfirm the user-defined alarm limit.
3. If not needed, disable this alert.
6.2.11
HI Alarm
The primary variable has surpassed the user-defined limit.
WARNING
Always respond to all indications of dangerous concentrations of toxic gas as though they
are real until they are proven false.
Alerts for more information.
Troubleshooting