Emerson Rosemount 928 Reference Manual page 48

Wireless gas monitor
Hide thumbs Also See for Rosemount 928:
Table of Contents

Advertisement

Configuration
August 2018
2.8.1
Calibrate using Field Communicator
Complete the following steps to calibrate the Rosemount
Field Communicator.
About this task
Note
Emerson
this manual using Emerson AMS Trex Device Communicator. The menus are identical to
those found in other Field Communicators, but you navigate using touch screens rather
than fast keys. Refer to the manual for your handheld communication device for more
information.
A. +COMM terminal
B. -COMM terminal
Procedure
1. Connect the HART
2. Establish communication between the Rosemount 928 Transmitter and the Field
3. On the Home screen, select Configure.
48
developed the Field Communicator guided setup configuration procedures in
®
communication leads from the Field Communicator HART
terminals to the COMM terminals on the terminal block of the Rosemount 928
Transmitter.
WARNING
Do not connect to the COMM terminals when an explosive atmosphere is present.
Communicator.
A B
Reference Manual
00809-0100-4928
628 Gas Sensor Module using a
Rosemount 928