5.6 Configuring the Hardware

Before the Ultium system can be used, the device software settings must be configured to recognize
the different components that make up the system. Follow the below instructions to update the
receiver firmware, sensor firmware, and populate sensors to prepare for a data collection.
Step 1
Open MR3, typically listed under Noraxon ->
MR3
Click on the Hardware Setup
Hardware Setup button in the upper
Hardware Setup
Hardware Setup
right hand corner.
Step 2
Select the Ultium icon, within the 'New Device
column, and click on the Insert
Note: Make sure the Receiver is attached to the
USB port of the computer.
Step 3
The Ultium Settings dialog will appear as shown.
Within the "Settings" tab, select the desired
collection Sampling Rate.
Select a RF Network from the RF Network list. In
most cases the default "1" will work. However, if
using multiple Ultium systems they must be
placed on separate networks.
If using MyoSync, check the Use Noraxon
MyoSync checkbox (see the Ultium EMG User
Manual for more information).
You may also set EMG filter cutoffs here.
(Rev J)
'New Device' ' ' '
'New Device
'New Device
Insert button.
Insert
Insert
17