Ultium EMG Hardware User Manual
Performance Characteristics
Output & Transmission Frequency (Depending on country)
Output & Transmission Frequency (Depending on country)
Output & Transmission Frequency (Depending on country)
Output & Transmission Frequency (Depending on country)
Up to 100 mW (depending on country allowance)
GFSK 2402-2480 MHz, frequency hopping
Utilizing up to 79 hopping channels
Channel-to-channel hopping frequency 50Hz (20 msec)
EMG sensor transmission range: 40m (typical)
EMG Sensor Data Acquisition System
EMG Sensor Data Acquisition System
EMG Sensor Data Acquisition System
EMG Sensor Data Acquisition System
16-bit resolution
Selectable Sample Rate: 2000Hz or 4000Hz
EMG Specifications
EMG Specifications
EMG Specifications
EMG Specifications
No notch (50/60 Hz) filters are used
Selectable high-pass cutoff: 5Hz, 10Hz or 20Hz
Selectable low-pass cutoff: 500Hz, 1000Hz or 1500Hz
Internal 24-bit sampling
Baseline noise < 1 uV RMS
Input impedance > 100 MOhm
CMR > 100 dB
Input range: +/- 24,000 uV
Adaptive Resolution:
< +/- 5,000uV: 0.3uV
o
> +/- 5,000uV: 1.16uV
o
Measurement Functional Accuracy: +/-1% F.S.
Sensor operation up to 8 hours on a fully charged battery (recharge time 3 hours)
Pinch-style terminal electrode connections
Energy Consumption, Condition of Use
EMG Sensor Charger is powered by 5VDC at 3.0A using a 110-240 VAC 50/60 Hz external
supply
Environmental Conditions for Normal Operation
(Rev J)
Ultium Charging Dock
Ultium Charging Dock
Ultium Charging Dock
Ultium Charging Dock
Dimensions
Dimensions
Dimensions
Dimensions
10.3" L x 1.4" W x 1.1" H
(26.1 cm L x 3.6 cm W x 2.9 cm H)
Weight
Weight
Weight
Weight
Less than 5.8oz (165g).