Noraxon Ultium EMG System User Manual page 39

Sensor and receiver
Hide thumbs Also See for Ultium EMG System:
Table of Contents

Advertisement

Ultium EMG Hardware User Manual
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic Emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
The Ultium System is intended for use in electromagnetic environment specified below. The customer or the user of the
Ultium System should assure that it is used in such an environment.
Immunity Test
IEC 60601
test level
Electrostatic
±8 kV contact
discharge (ESD)
± 15 kV air
IEC 64000-4-2
Electrical fast
±2kV for power supply
transient/burst
lines
IEC 61000-4-4
±1kV for input/output
lines
Surge
±1kV differential
mode
IEC 61000-4-5
±2kV common mode
Voltage dips, short
<5 % U
interruptions and
(>95 % dip in U
voltage variations
for 0,5 cycles
on power supply
input lines
40 % U
(60 % dip in U
IEC 61000-4-11
for 5 cycles
70 % U
(30 % dip in U
For 25 cycles
<5 % U
(>95 % dip in U
For 5 secs
Power frequency
3 A/m
(50/60 Hz) magnetic
field
IEC 61000-4-8
NOTE
U
is the a.c. mains voltage prior to application of the test level.
T
Guidance and manufacturer's declaration – electromagnetic immunity
The Ultium System is intended for use in electromagnetic environment specified below. The customer or the user of the
Ultium System should assure that it is used in such an environment.
Immunity Test
IEC 60601 test level
(Rev J)
The Ultium System must emit electromagnetic energy in
Group 2
order to perform its intended function. Nearby electronic
equipment may be affected.
Class A
The Ultium System is suitable for use in all establishments
other than domestic establishments and those directly
connected to the public low-voltage power supply network
Not applicable
that supplies buildings used for domestic purposes.
Not applicable
Compliance level
±8 kV contact
±15 kV air
±2kV for power supply
lines
Not applicable
±1kV differential mode
±2kV common mode
Not applicable to
T
)
operation
T
Not applicable to
T
)
operation
T
Not applicable to
T
)
operation
T
Not applicable to
T
)
operation
T
3 A/m
Compliance
level
Electromagnetic environment -
guidance
Device user should avoid touching subject
and sensor probes while a measurement
is active.
For battery charging mains power quality
should be that of a typical commercial or
hospital environment.
For battery charging mains power quality
should be that of a typical commercial or
hospital environment.
For battery charging mains power quality
should be that of a typical commercial or
hospital environment.
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or hospital
environment.
Electromagnetic environment -
guidance
39