Additional options
8.1 Automatic calibration
8.1.1 One point calibration with a volumetric pump
Figure 11 Automatic calibration option
1 Pump
2 Elbow fitting
38
This option allows the analyzer to perform an automatic calibration at programmed
intervals and with a known calibration solution. Three different kits are available, each
including a 10 liter canister equipped with level detector and strainer:
1. 1 point calibration with a volumetric pump.
2. 2 point calibration with a micro piston pump.
3. 2 point calibration with a pulse pump.
For information regarding the programming and use of this option, refer to the section
entitled
Calibration (CALIBRATION option)
Electrocution hazard. Always make sure the analyzer is turned off and power is disconnected
before installing this option.
3 Polyethylene tubing
4 Connection to attach to top of input
block
Install the pump as follows (the numbers in brackets refer to
1. Mount the pump (1) in position 4 on the rear of the panel (see
2. Connect the tubing (3) between the pump (2) and the input block (5) installed on the
measurement cell cover using the fitting (4).
3. Connect the cable (6) to terminals 1 and 11 of J14 in the electronic unit, and make
sure that jumper X4 is installed between terminals 2 and 3 and a strap placed
between terminals 8 and 10 of J14 (refer also to
on page 20).
To install the 10 liter canister, refer to
on page 26.