Product overview
3.2 Measurement chamber
Figure 4 Measurement chamber components
1 Stirrer motor
2 Measurement chamber cover
3 Electrode
3.3 Features
10
4 Electrode cable
5 Siphon
6 Drain valve
The measurement chamber has mounting locations for:
•
Selective electrode: Measurement of potential.
•
Combined electrode: Single combined selective and reference electrode.
•
Reference electrode: Measurement of potential.
•
Pt 100: Temperature measurement.
•
Sprinkler: Uses rinse water to clean the measurement chamber between
measurements.
•
Chemical cleaning (option): The cleaning can be improved with the addition of a
chemical reagent to the rinse water.
•
Siphon: Used to ensure a constant and reproducible volume of sample in the
measurement chamber.
•
Overflow pipe: To avoid any sample overflow from the measurement chamber.
•
Reagent additions: Reagents and calibration solutions are added to the measurement
chamber by peristaltic or micro piston pumps.
•
Stirrer: Mixes the liquids present in the measuring cell.
•
Dilution system (option): Used to dilute samples of high concentration.
•
Heating system (option): Used to heat the sample to a defined temperature before
measurement.
7 Drain tubing
8 Overflow pipe
9 Threaded fittings