Section 4

4.1 Unpacking

4.2 Mechanical installation

4.2.1 Mounting the analyzer

4.2.1.1

Panel mounting

4.2.1.2

Wall mounting

Installation
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document. Mains power should only be connected once installation has been completed and
checked.
Carefully remove the analyzer and its accessories from the box and packing material,
referring to the packing list included to confirm that everything has been delivered. Please
visually inspect the analyzer for shipping damage. If anything is missing or damaged,
contact the manufacturer or your dealer immediately.
You may want to retain the box and other packing material in case later you need to ship
the analyzer. Please dispose safely and ecologically of the box and packing material (if
not stored for future use).
Please read through this manual thoroughly before carrying out the installation.
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
The analyzer should be mounted as near as possible to the sample inlet, and should be located in
an easily accessible area to facilitate periodic checking of the sample flow rate, and for regular
maintenance. Wherever the analyzer is to be mounted, it is important to note that it must be
placed in an upright position with the electronic unit (No. 1 in
recommended to use a spirit level to ensure that the analyzer is correctly positioned and not
leaning to one side or forward. This is essential to guarantee the accuracy of the analyzer.
This model is designed as a 19 inch (48.26 cm) rack system. Delivery includes six
M6 screws for mounting the panel onto a rack. All internal connections are factory
prepared. Canisters for reagents can be fitted onto an optional storage tray.
All internal connections are factory prepared. Cables and tubing are threaded through the
cable glands located on the bottom right side of the cabinet. Canisters for reagents can
be fitted onto an optional storage tray.
To open the cabinet, push the handle lid upwards, press the lock button, then open the
door with a 45° left turn of the handle. To access the cabinet interior, remove the knurled
screw on the right side of the panel and carefully turn the panel sideways to the left taking
care not to squeeze any tubing.
D A N G E R
C A U T I O N
N O T I C E
Figure 3
on page 9) at the top. It is