8.8.3 Pulse pump

Figure 19 Pulse pump

8.9 Canister kit

48
®
4. Connect Tygon
tubing from the pump (2) to the measuring cell.
®
5. Connect Tygon
tubing from the pump (1) to the reagent canister.
6. Connect the pump cable (4) to the electronic unit according to the pump function
(refer to
Table 6
on page 20 and
7. Reconnect power to the analyzer.
8. Adjust the flow rate as required (3).
Electrocution hazard. Always make sure the analyzer is turned off and power is disconnected
before installing this option.
Install the pump as follows:
1. Install the pump on the front of the panel in one of the available slots (see
on page 37) using the 2 screws.
2. Connect the pump cable (3) to the electronic unit according to the pump function
(refer to
Table 6
on page 20 and
3. Replace the cover on the back of the panel using the 4 screws.
4. Connect the outlet OUT (1) to the tube (4) using Tygon
into one of the five slots in the measurement chamber cover.
5. Connect the inlet IN (2) to the reagent canister using Tygon
Electrocution hazard. Always make sure the analyzer is turned off and power is disconnected
before installing this option.
The reagent pumps, automatic calibration and the chemical cleaning options can all be
delivered with a 10 liter canister equipped with a level detector, strainer, and tubing for
connection to the pumps.
Refer to
Reagent level detector
the analyzer.
Table 7
on page 20).
D A N G E R
Table 7
on page 20).
D A N G E R
on page 20 for information on wiring the level detector to
Figure 10
®
1.6 x 3.2 tubing and insert
®
1.6 x 3.2 tubing.