Additional options
Figure 14 Dilution reservoir
1 Elbow fitting
2 Reservoir stopper
3 Hydraulic plunger
4 Connector opening
42
Note: For optimal performance, the minimum rinse water pressure should be 2.5 bar.
Install the dilution assembly as follows (the numbers in brackets refer to
1. Install the control valve on the panel (position 4 in
2. Remove the overflow pipe (position 8 in
assembly (13) in its place in the measurement cell using the Teflon washer and O-
ring.
3. Install the PE tubing 3.2/6.4 (14) between the control valve (12) and output A of the
dilution reservoir.
4. Install the PE tubing 3.2/6.4 (15) between the control valve (12) and output B of the
dilution reservoir.
5. Connect the cable (16) to terminals 6 and 7 of J14 in the electronic unit (refer also to
Figure 8
on page 18 and
6. Connect the compressed air between the outlet (19) and inlet (20) connections.
7. Adjust the opening and closing pressures of the system using the adjustment screw
(17). The system should close as slowly as possible.
8. Connect the sample to the input (10) with 12 mm internal diameter tubing (pressure
0.5 - 6 bar)
9. Connect the rinse water to the input (11) with 6 mm internal diameter tubing (pressure
2.5 - 6 bar)
10. Connect input P on the T connection (18) to the compressed air supply (pressure 4 -
7 bar)
Dilution reservoir
5 Locking hole
6 Plunger
7 Reservoir
8 Gasket
Disassemble the reservoir as follows (the numbers in brackets refer to
1. Remove the connector (1) by unscrewing it from the reservoir.
2. Remove the stopper (2).
3. Pull and turn the plunger (3) until you see the locking hole (5) through the connector
opening (4).
4. Block the plunger (6) through the openings (4) and (5).
5. Unscrew the plunger (3).
6. Remove the plunger (6) by pushing on it.
7. Clean the two plungers (3) and (6).
8. Clean the inside of the reservoir (7) using a soft non-abrasive cloth.
9. Lubricate the gaskets (8) and (9) or change them if necessary.
Figure 4
Table 6
on page 20).
Figure
Figure 10
on page 37).
on page 10) and install the dilution
9 Gasket
10 O-ring
11 O-ring
Figure
13):
14):