Using the 8990B

Accessing the Multipulse Mode

Multipulse is only available in the RF channels with a maximum of 512 frames. It
can be used to analyze the power amplifier or transmitter output stability by
measuring the amplitude droop, pulse to pulse stability, and abnormal pulse.
Multipulse is also able to capture and analyze continuous pulses, compare the
pulses characteristics, and capture long PRI pulses or any multiple pulse trains or
burst. The histogram in the multipulse mode is useful to analyze the jitter in rise
time, fall time, and pulse width measurements of the pulse trains.
Use the following procedure to display the multipulse mode.
1 Select Acquisition > MultiPulse on the toolbar as shown in
Figure 2-90
2 The MultiPulse Setup dialog is displayed as shown in
Figure 2-91
120
MultiPulse selection in the Acquisition menu
to configure the multipulse display.
MultiPulse Setup dialog
Figure
2-90.
Figure 2-91
to allow you