Invalid character data
141
Either the character data element contains an invalid character, or the element is not valid
Character data too long
144
The character data element contains more than 12 characters
Character data not allowed
148
A discrete parameter was received but a character string or a numeric parameter was expected. Check the
list of parameters to verify that you have used a valid parameter type.
Example: MARK:MODE OFF_0
String data error
150
Generic string data error
Invalid string data
151
An invalid string was received. Check to see if you have enclosed the character string in single or double
quotes.
Example; SYST:COMM:LAN:ADDR "10.0.0.2
String data not allowed
158
A character string was received but is not allowed for the command. Check the list of parameters to verify
that you have used a valid parameter type.
Example: SYST:LOCK 'ON'
Block data error
160
Generic block data error
Invalid block data
161
A block data element was expected but was invalid for some reason
Block data not allowed
168
A legal block data element was encountered but not allowed by the 8990B at this point
Expression error
170
Generic expression error
Invalid expression data
171
The expression data element was invalid
Expression data not allowed
178
A legal expression data was encountered but not allowed by the 8990B at this point
Example: MARK:LEV (5+5)
Execution error
200
Generic syntax error
Keysight 8990B User's Guide
Maintenance
3
183