Error List - Keysight Technologies 8990B User Manual

Peak power analyzer
Table of Contents

Advertisement

Error list

0
No error
The error queue is completely empty. Every error or event in the queue has been read or the queue has been
purposely cleared by power-on, *CLS, and so forth.
Command error
100
Generic syntax error
Invalid character
101
An invalid character was found in the command string. You may have inserted a character such as #, $, or %
in the command header or within a parameter.
Example: MARK:AXP 30#
Syntax error
102
An invalid syntax was found in the command string
Example: AUTO
Invalid separator
103
An invalid separator was found in the command string. You may have used a comma instead of a colon,
semicolon, or blank space; or you may have used a blank space instead of a comma.
Example: ACQ:AVER:COUN,128
GET not allowed
105
A Group Execute Trigger (GET) is not allowed within a command string
Missing parameter
109
Fewer parameters were received than expected for the command. You may have omitted one or more
parameters that are required for this command.
Example: CHAN1:BWID
Command header error
110
An error was detected in the header
Header separator error
111
A character that was not a valid header separator was found in the command string
Program mnemonic too long
112
A command header was received which contained more than the maximum 12 characters allowed
Example: SYSTemCOMMunicateLANDGATeway "10.0.0.2"
Keysight 8990B User's Guide