Benning CM 8 Operating Manual page 27

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 20
See figure 3:
8.3 Current measurements (frequency measurements, THD-harmonic distortion)
8.3.1
Preparing the measurement
Operate and store the BENNING CM 8 at the specified storage and operating
temperatures only! Do not permanently expose the device to sunlight.
-
Strong sources of interference in the vicinity of the BENNING CM 8 might
involve unstable readings and measuring errors.
Do not apply any voltage to the output contacts of the
BENNING CM 8! Any possibly connected safety measuring
leads have to be removed.
8.3.2
Current measurements
-
Select the desired function (A AC/ DC) by means of the rotary switch 8 of
the BENNING CM 8.
-
Select the current type (direct current DC or alternating current AC) to be
measured by means of the blue key 5 of the BENNING CM 8 and perform
a null balance N, if necessary.
-
Operate the opening lever K and clamp the single-wire live conductor cen-
trally by means of the clamp M of the BENNING CM 8.
-
Read the value indicated on the digital display 1.
-
In the AC range, it is possible to switch over to the frequency measuring mode
by pressing the „PEAK / Hz-THD" key 7 for approx. 2 seconds. You can
switch back by pressing the key for another 2 seconds. Normal pressing of
the key switches the device into the THD mode (%harmonic distortion).
See figure 4:
8.4 Inrush current measurements (AC)
-
Select the function (A) by means of the rotary switch 8 of the
BENNING CM 8.
-
Select the current type AC current by means of the blue key 5 of the
BENNING CM 8.
-
Briefly press the „INRUSH" key 6.
-
Operate the opening lever K and clamp the single-wire live conductor cen-
trally by means of the clamp M of the BENNING CM 8.
-
Switch on the load.
-
Read the value indicated on the digital display 1.
See figure 4:
8.5 Resistance measurements
-
Select the function (Ω) by means of the rotary switch 8 of the
BENNING CM 8.
-
Connect the black safety measuring lead to the COM jack J of the
BENNING CM 8.
-
Connect the red safety measuring lead to the jack for V and Ω 9 of the
BENNING CM 8.
-
Bring the safety measuring leads into contact with the measuring points and
read the measured value on the digital display 1 of the BENNING CM 8.
See figure 5:
8.6 Diode tests
-
Select the function (
BENNING CM 8.
-
Connect the black safety measuring lead to the COM jack J of the
BENNING CM 8.
-
Connect the red safety measuring lead to the jack for V, Ω 9 of the
BENNING CM 8.
-
Bring the safety measuring leads into contact with the diode connections and
read the measured value on the digital display 1 of the BENNING CM 8.
-
For a standard Si diode applied in conduction direction, a conduction volt-
age between 0.400 V and 0.900 V is displayed. „000" indicates a short-
circuit inside the diode, „OL" indicates an interruption inside the diode.
-
For a diode applied in reverse direction, „OL" is indicated. If the diode is
defective, „000" or other values are indicated.
See figure 5:
02/ 2019
AC voltage measurement/ frequency measurement
DC/ AC current measurement
(frequency, inrush current measurement)
DC/ AC current measurement
(frequency, inrush current measurement)
Resistance measurement/ diode test/
continuity test with buzzer
) by means of the rotary switch 8 of the
Resistance measurement/ diode test/
continuity test with buzzer
BENNING CM 8
19

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents