Table C-2
Problem
Vertex Monitoring
Vertex Monitoring does not work.
Vertex Monitoring alarm beep goes off too often.
Pupil Size
Not being automatically displayed on printout or on
screen.
Liquid Trial Lens
Error messages occurring.
External Keyboard
Does not work.
Trackpad, Trackball or Mouse
Does not work.
SWAP (Short-Wavelength Automated Perimetry)
SITA-SWAP not available with Blue-Yellow switched
to on.
Grayscale looks very dark.
Blind spot monitor reports many Fixation Losses.
Blue stimulus is on but bowl is not yellow.
Illumination from yellow lamp at top of bowl
annoying to patient.
Testing Problems
Test completes immediately after Start button is
pressed.
Full field test stops before complete.
Hard Disk Problems
Message indicates hard disk failure.
Patient names exist in database without test data.
Go to Contents
Possible Cause or Solution
• HFA3 not Models 850 or 860.
• Vertex Monitoring not turned on.
• Trial lens holder not in up position.
• Gaze Tracking was not initialized at the start of the test.
• Patient moving away from trial lens - reposition patient.
• Vertex Monitoring needs to be reset.
• Turn off Vertex Monitoring.
• HFA3 not Models 840, 850, or 860.
• Gaze Tracking was not initialized.
• Auto pupil turned off.
• Do not move the Liquid Trial Lens when it is being adjusted (icon flashing red).
- To resolve error messages, lower then raise the lens. If errors persist, remove the Liquid
Trial Lens holder completely and then reattach it.
• May not be compatible with HFA3.
- Try another keyboard.
• Not plugged in properly.
• May not be compatible with HFA3.
- Try a different trackball/mouse.
• Wrong test pattern used.
- Choose a 24-2 Threshold test, then turn Blue-Yellow on.
• May be normal for Blue-Yellow. The eye is less sensitive to the blue stimulus. 
- Gray scale is the same one used for White-on-White.
- Use Total and Pattern Deviation plots for diagnosis.
• Patient had difficult time recognizing when to respond.
- Re-instruct the patient and repeat the test.
• Size V blind spot check size may be too easy to see at blind spot.
- Consider using Gaze Track only and turn off Blind Spot.
• Blue stimulus color was chosen instead of Blue-Yellow parameter.
- Blue Yellow parameter set to On will give yellow bowl and Blue Size V stimulus.
• Visor not extended during testing procedure.
• Simulation Mode is on.
- Turn Simulation Mode off.
• Trial lens holder in up position for peripheral portion of test.
- Test peripheral field without a trial lens and put holder down.
• Call Customer Service and mention the error code.
• Test results not saved with patient information.
- Delete patient record.
HFA3 Instructions for Use
Troubleshooting
2660021166131 Rev. A 2018-11