Code
Meaning
/Icon
S: 31
Flash error.
S: 33
Factory calibration for the internal tem-
perature sensor has been lost.
S: 34
Factory calibration for the device has
been lost.
S:39
Transducer temperature sensor failure.
S:40
Transducer temperature too high.
S: 41
Transducer temperature too low.
S:64
Device error.
NOTE: Fault text and icon appear only
on LCD.
S:66 to
Device error.
S:83
NOTE: Fault text and icons appear only
on LCD.
7ML19985LY01
General Fault Codes (cont'd)
SITRANS LR560 (FF) – INSTRUCTION MANUAL
Corrective Action
Repair required: contact your local
Siemens representative.
Repair required: contact your local
Siemens representative.
Repair required: contact your local
Siemens representative.
Repair required: contact your local
Siemens representative.
Relocate device and/or lower
process temperature enough to cool
device. Inspect for heat-related
damage and contact your local
Siemens representative if repair is
required.
Relocate device and/or raise process
temperature enough to warm device.
Inspect for temperature-related
damage and contact your local
Siemens representative if repair is
required.
Repair required. Contact your local
Siemens representative.
Repair required. Contact your local
Siemens representative.
Page 133